Detailed Reference Information |
Holdren, D.H., Olsen, R.O. and Schmidlin, F.J. (2001). Comparison of total ozone overburden from handheld photometers with the Wallops Island Dobson Spectrophotometer. Geophysical Research Letters 28: doi: 10.1029/2000GL011517. issn: 0094-8276. |
|
Total ozone column measurements are compared between the Dobson spectrophotometer located at Wallops Flight Facility, Wallops Island, VA, and two Microtops II photometers. The purpose of these comparisons is to determine whether the Microtops II ozone measurements are comparable to those obtained with the Dobson spectrophotometer. Results have shown that the handheld photometer can provide measurements within about 2% of the Dobson's. A comparison between the two Microtops showed approximately a 3% difference in the derived ozone overburdens. Better results with the two Microtops were obtained using the 305.5-nm and the 312.5-nm filters only. ¿ 2001 American Geophysical Union |
|
 |
 |
BACKGROUND DATA FILES |
|
 |
Abstract |
|
 |
|
|
|
Keywords
Atmospheric Composition and Structure, Instruments and techniques, Global Change, Instruments and techniques |
|
Publisher
American Geophysical Union 2000 Florida Avenue N.W. Washington, D.C. 20009-1277 USA 1-202-462-6900 1-202-328-0566 service@agu.org |
|
|
 |